Speed up separations and save solvents with Merck UHPLC columns

Merck UHPLC columns

The use of columns with a 2.1 mm inner diameter will reduce solvent consumption by up to 90%. In addition, by using a UHPLC column from Merck, packed with the reversed-phase sorbent Purospher® STAR RP-18 endcapped 2 µm, the separation time can be decreased by a factor of 10.

In typical routine HPLC laboratories, standard columns of 4.6 mm internal diameter (I.D.) are still used for most applications. On average, such laboratories require 60 L of solvent per year. The used solvents are mostly acetonitrile, methanol, isopropanol, hexane, and/or aqueous buffers. As a result, they generate 60 L of mixed organic/aqueous waste each year.

Purospher® STAR RP-18 endcapped UHPLC columns are ideal for ultra-fast applications where resolution, sensitivity, and sample throughput are crucial. They are the first choice for high-throughput screening and quality control analyses, process monitoring, method development, and LC/MS applications. Due to its balanced selectivity, a Purospher® STAR RP-18 endcapped UHPLC column can be used for a broad range of demanding separations with tailing-free chromatograms.

 

Purospher® STAR RP-18 endcapped UHPLC

By using a 50-2.1 UHPLC column packed with Purospher® STAR instead of a 150-4.6 column packed with Purospher®, for the separation of alkylphenones, solvent consumption can be reduced by 89%. Additionally, the separation time can be decreased by a factor of 10.

 

Separation of alkylphenonesSample: Alkylphenone standard
1. Acetanilide
2. Acetophenone
3. Propiophenone
4. Butyrophenone
5. Benzophenone
6. Valerophenone
7. Hexanophenone
8. Heptanophenone
9. Octanophenone
Purospher® STAR RP-18 endcapped (5 µm), 150-4.6 mm
Column Purospher® STAR RP-18 endcapped (5 µm), 150-4.6 mm Separation with Purospher® STAR RP-18 endcapped (5 µm), 150-4.6 mm
Eluent A: water
B: acetonitrile
Gradient

0.0 min
15.0 min
15.1 - 20 min with reequilibration

45% B       
45 - 95% B
45% B
Flow rate   1.0 mL/min
Pressure 105 bar
Detection 247 nm
Temperature 40°C
Inj. volume 10 µL
Total run time 20 min
Solvent consumption per separation 20 mL
Purospher® STAR RP-18 endcapped (2 µm), 50-2.1 mm
Column Purospher® STAR RP-18 endcapped (2 µm), 50-2.1 mm Separation with Purospher® STAR RP-18 endcapped (2 µm), 50-2.1 mm
Eluent A: water
B: acetonitrile
Gradient

0.0 min
0.8 min
0.9 - 2 min with reequilibration

55% B         
55 - 100% B
55% B
Flow rate   1.1 mL/min
Pressure 505 bar
Detection 247 nm
Temperature 40°C
Inj. volume 1 µL
Total run time 20 min
Solvent consumption per separation 2.2 mL

Purospher® STAR UHPLC columns are available in different dimensions, packed with 2 µm or 3 µm particles. The Purospher® STAR RP-18 endcapped 3 µm columns are recommended for difficult samples where clogging and high backpressure is an issue.

Ordering
Number
DesignationColumn
dimension
Content of
pack
1506450001 Purospher® STAR RP-18 endcapped 2 µm 30-2.1 1 UHPLC column
1506460001 Purospher® STAR RP-18 endcapped 2 µm 50-2.1 1 UHPLC column
1506480001 Purospher® STAR RP-18 endcapped 2 µm 100-2.1 1 UHPLC column
1506490001 Purospher® STAR RP-18 endcapped 2 µm 150-2.1 1 UHPLC column
       
1506500001 Purospher® STAR RP-18 endcapped 3 µm 30-2.1 1 UHPLC column
1506510001 Purospher® STAR RP-18 endcapped 3 µm 50-2.1 1 UHPLC column
1506530001 Purospher® STAR RP-18 endcapped 3 µm 100-2.1 1 UHPLC column

© 德国默克,达姆施塔特,德国 2012

x

欢迎来到默克化工

 

欧洲,中东和非洲 亚洲 / 太平洋地区 北美,美国中部和加勒比 南美

请选择国家

北美,美国中部和加勒比

南美

欧洲,中东和非洲

亚洲 / 太平洋地区

A 到 Z的所有国家

如果不能找到你所在的国家,请点击此处 此处.

 

重要信息


目前无法购买生化产品以及默克的其他产品。我们建议您先完成目前的订单。

是否继续操作 没有 完成您目前的订单? 如果是,您的购物车将清空。